Cost-effective 12ch ECG
CardioCare
2000
The CardioCare is a 12 channel resting ECG which
is the most economically priced without performance compromise,
especially suitable for routine use in Private check, ER or Hospitals.
The design of CardioCare is maximized for the user's
convenience (A4 size paper, one body type). One of special characteristics
is the upgrade and option which makes it possible to be connected
in various types of network system: Fax, PC(RS-232 port or LAN).
|
|
|
Specifications
| ECG leads |
12 leads simultaneous 10-second resting
ECG |
| Dimension |
296(W) x 305.5(H) x 92.5(D)mm, Approx.
2.98 kg |
| Recording channel |
3,6,12 channel and 1 channel (60 sec) |
| Sensitivity |
5,10,20, auto(I~aVF:10,V1~V6:5) mm/mV |
| Printing speed |
12.5, 25, 50 mm/s |
| Filters |
AC(50/60 Hz, -20 dB or better)
Muscle (25 ~ 35 Hz, -3dB or better)
Base line drift (0.1 Hz, -3dB or better)
Low pass filter (off, 40 Hz, 100 Hz, 150 Hz) |
Display
LCD display
LED indicator |
2 x 16 character LCD Display
Sensitvity, speed, filter status, HR, printing form, rhythm lead
signal quality, power source, battery. |
| Keyboard |
Membrane keyboard, Alphanumeric and
symbol available |
Patient data
Basic measurement |
ID, name, age, sex, height, weight
Heart rate, PR, QRS, QT/QTc, P-R-T axis |
| Recorder |
Thermal print head, Thermal roll paper
Report papers :
width : A4 : 210mm or 8.5"
length : A4 : 300mm or 11"
Effective writing width : 204 mm
Resolution :
Vertical: 8 dots/mm
Horizontal : 16dots/mm |
| Electrical |
Internal noise: 20uV (p-p) max
Input circuit: Floating input
Isolated and defibrillation protected
Input impedance: =10M Q
Input range: =±5mV
Common Mode Rejection: = +100dB
DC offset voltage: >±300mV
Time constant; 3.2sec
Patient leakage current: <10µA
Frequency response 0.05-150 Hz |
| Signal quality control |
Disconnected lead detection |
| Power |
Power supply: AC or built-in Battery(option)
Power requirement: 100 - 240 Vac, 50/60Hz, 1.0 - 0,5 A
Power consumption: 60W max |
| Battery capacity |
1 hour of normal use(approximately
100 automatic ECG printouts)
|
| Communication |
PC connection with RS-232 interface
and LAN |
Safety
Conformity |
Class I, Type BF
CE,CSA |
| Environment |
Operating humidity: 30 ~ 85%
Operating Temperature: 10° ~ 40°C
Atmospheric pressure: 70 ~ 106 KPa |
| Standard accessory |
Power Cable 1ea, Patient Cable 1ea,
Electrode 10ea, Chart Paper 1ea, Operation Manual 1ea |
| Options |
Rechargeable Battery(NI-MH), Cart,
Hanger, Bag, PC s/w, Limb electrode, Chest electrode, RS-232C null
cable, LAN cable(Ethernet-RJ-45) |
|
|